Separate caps for performing filtration and sparging/delivery
· Filtration/vacuum degassing can be performed at one central location with prepared mobile phase delivered directly to HPLC stations
· Filtration cap is specially designed to filter HPLC mobile phase directly into the reservoir using 47 mm membrane filters
· Vacuum degassing is accomplished by closing the HI-VAC® valve on the solvent pickup adapter
· Three hole cap is used to helium sparge and deliver the mobile phase to the pump system
· Helium-tight seals maximize sparging efficiency
· 1/4-28 fittings and adapters are supplied to allow simple, easy connection to any HPLC pump system
· Supplied with filtration cap (with 40-60 micron porosity glass support frit), standard solvent pickup adapter, three hole cap, conical bottom reservoir, standard fittings kit and FEP PTFE tubing
· Patent No. 5,397,467
· Recommended for most solvents used for HPLC mobile phase
· Caps and solvent pickup adapter o-rings and solvent inlet filters must be changed when using tetrahydrofuran (THF) or other aggressive solutions
· Manufactured from 33 expansion, low extractable borosilicate glass conforming to USP Type I and ASTM E438, Type I, Class A requirements